Demonstration of magnetically silent optically pumped magnetometers for the TUCAN electric dipole moment experiment

dc.contributor.authorKlassen, Wolfgang
dc.contributor.authorAhmed, Shomi
dc.contributor.authorGrehan, Kiera P.
dc.contributor.authorHovde, Chris
dc.contributor.authorMadison, Kirk W.
dc.contributor.authorMammei, Russell R.
dc.contributor.authorMartin, Jeffery W.
dc.contributor.authorMcCrea, Mark
dc.contributor.authorMohammadi, Tahereh
dc.contributor.authorMomose, Takamasa
dc.contributor.authorOpsahl, Patrick
dc.contributor.authorOstapchuk, David C. M.
dc.date.accessioned2024-12-06T20:36:34Z
dc.date.available2024-12-06T20:36:34Z
dc.date.issued2024-11-19
dc.date.updated2024-12-01T04:17:23Z
dc.description.abstractAbstract We report the performance of a magnetically silent optically pumped cesium magnetometer with a statistical sensitivity of 3.5 pT/ Hz at 1 Hz and a stability of 90 fT over 150 s of measurement. Optical pumping with coherent, linearly-polarized, resonant light leads to a relatively long-lived polarized ground state of the cesium vapour contained in a measurement cell. The state precesses at its Larmor frequency in the magnetic field to be measured. Nonlinear magneto-optical rotation then leads to the rotation of the plane of polarization of a linearly polarized probe laser beam. The rotation angle is modulated at twice the Larmor frequency. A measurement of this frequency constitutes an absolute measurement of the magnetic field magnitude. Featuring purely optical operation, non-magnetic construction, low noise floor, and high stability, this sensor will be used for the upcoming TUCAN electric dipole moment experiment and other highly sensitive magnetic applications. Novel aspects of the system include commercial construction and the ability to operate up to 24 sensors on a single probe laser diode.
dc.identifier.citationThe European Physical Journal C. 2024 Nov 19;84(11):1181
dc.identifier.doi10.1140/epjc/s10052-024-13544-5
dc.identifier.urihttp://hdl.handle.net/1993/38697
dc.language.isoeng
dc.language.rfc3066en
dc.publisherSpringer Open
dc.rightsopen accessen_US
dc.rights.holderThe Author(s)
dc.titleDemonstration of magnetically silent optically pumped magnetometers for the TUCAN electric dipole moment experiment
dc.typeJournal Article
local.author.affiliationFaculty of Science::Department of Physics and Astronomy
oaire.citation.issue1181
oaire.citation.titleThe European Physical Journal C
oaire.citation.volume84
Files
Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
10052_2024_Article_13544.pdf
Size:
1.26 MB
Format:
Adobe Portable Document Format
Description:
License bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
license.txt
Size:
770 B
Format:
Item-specific license agreed to upon submission
Description: